The LBS pipette satisfies the highest requirements for accuracy and precision while offering the best ergonomics, more flexibility, and a smart combination of features and usefulness.
It is simple to calibrate and maintain with the tools provided.
LBS provides both single channel and multi-channel pipette with precision for research works
These LBS 8-channel pipettes are suitable for standard 96-well plates.
The LBS pipettes come with an operator manual that explains how to recalibrate and maintain it.
Product Specification
Lightweight and Ergonomic design provides excellent operating, very suitable for gripping. Light spring, press very labor-saving, help to relieve fatigue of hand joints, easy suction.
Digital display, adjusted delivery volume clearly indicated in the large digital display on the handle.
Fully made of high strength material. Anti-strengthening corrosion, alcohol disinfection; Enhanced anti UV linearity, UV sterilization;
Different colors are used to identify pressing handles, making it easier to select pipettes of required specifications;
Well-compatible with most of the pipette tips on the market;
The pipette is equipped with universal pipette tips to ensure good air tightness and accurate pipetting volume.
Each LBS pipette is calibrated according to EN/ISO8655 standard, with strict quality control to ensure that each pipette is a high-quality product
The Luminous Single and multi channel pipettors fits most popular pipttors without any loss of accuracy or precision
Our Advantages
Light Weight: Lightweight spring for effortless pressing
The Best Quality: The pipettes are made from high strength material and anti-strengthening corrosion
Ensured Precision: They LBS pipettes have highest accuracy and precision.
Compatibility: Perfectly compatible with most pipette tips on the market
Pipette Operation
Forward Technique Pipetting (low density liquids)
Press the push button to the first STOP.
Slowly release the button to the ready position to complete the aspiration of the liquid;
Deliver the liquid by gently depressing the push button to the first STOP. After a delay of about one second, continue to depress the push button all the way to the second STOP. This action will empty the tip.
Release the push button to the ready position. If necessary, change the tip and continue pipetting.
Reverse Technique Pipetting (for high viscosity or volatile liquids)
Depress the push button all the way to the second STOP for aspirating liquid, slowly release the button to the ready postion to complete the aspiration of the liquid, aspirating more than the preset volume
Press the button to the first STOP and discharge the preset volume of liquid
The remaining liquid should be discarded with the tip.
Caution
Before liquid handling, make sure that the pipette and tips are at the same temperature, hold the pipette in an upright position while aspirating the liquid, and dip the tip under the surface of the liquid to a depth of 2~3mm. Fill and empty the tip by several times to moisten the tip nozzle before pipetting (particularly when working with high-viscosity liquids or density different from water).
If the pipette tip is full of liquid, do not turn the pipette upside down or lay it flat to prevent liquid from flowing into the cavity. It is recommended to use a special pipette stand when placing the pipette;
Make sure to use it within the volume range when pipetting;
If it is not in use for a long time, it is recommended to apply special maintenance oil to the sealing ring;
If need to disassemble, please find a professional technician for maintenance.
Calibration and Adjustment:
Take out the Service tool and place the Service tool as shown in the figure below, and turn the service tool clockwise to increase, or counterclockwise to decrease the volume. Calibrate pipettes 1-2 times per year (depending on frequency of use)